AS-717

Coliform & E. coli Chromogenic Contact Plates 55 mm | AS-717

$69.00 AUD ex GST
Number
Request a Quote
Ready-to-use Contact Plates · Chromogenic

Coliform / E. coli Chromogenic Contact Plates

Ready-to-use 55 mm chromogenic contact plates for detecting and differentiating coliforms and E. coli on surfaces, on personnel and in air.

Cat. No. AS-71755 mm · 25 cm²pH 7.2 ± 0.2Ready to useResults in 18–24 h
No preparation needed: plates are poured, sealed and ready to use. Press, incubate and read colonies by colour.

Colony Colours

E. coliBlue-green
Other coliformsMagenta / purple
Other organismsColourless
Incubation35–37 °C

Intended Use

Ready-to-use 55 mm chromogenic contact plates for detecting and differentiating coliforms and E. coli on surfaces, on personnel and in air. For hygiene and environmental monitoring programs in food, pharmaceutical and industrial facilities.

For laboratory microbiological testing, quality control and research use only. Not intended for direct use in food, human or veterinary therapeutic applications, diagnosis, or consumption.

Principle

Peptone and yeast extract supply nitrogen, carbon, vitamins and growth factors. Sodium dodecyl sulfate inhibits most Gram-positive organisms.

Coliforms produce β-D-galactosidase, which cleaves a chromogenic substrate in the medium and gives magenta to purple colonies. E. coli is detected by a second, E. coli-specific chromogenic substrate and forms blue-green colonies that are clearly distinguished from other coliforms.

Formulation (g/L)

Component g/L
Meat peptone 10.00
Yeast extract 7.00
Sodium chloride 5.00
Sodium dodecyl sulfate (SDS) 0.20
Chromogenic substrates (proprietary blend) 0.12
Agar 14.50
Total 36.82

Selective agent shown in amber. Formulation may be adjusted to meet performance specifications.

Key Applications

Food & BeverageHygiene monitoring of food-contact surfaces, equipment and packaging lines.
PharmaceuticalSurface and personnel monitoring in production and support areas.
Water FacilitiesHygiene checks of treatment, storage and distribution surfaces.
Oil & GasSurface monitoring around water systems, tanks and pipework.
Personnel & GlovesGlove-print and garment sampling for hand-hygiene programs.
Air (Settle Plate)Open-plate exposure, for example 15 minutes, in production rooms.

Reading the Results

Organism group Colony colour
Escherichia coli Blue-green
Other coliforms (Citrobacter, Enterobacter, Klebsiella, Serratia) Magenta / purple
Other Gram-negative organisms Colourless

Total coliforms = blue-green + magenta/purple colonies. Report as CFU per plate (CFU/25 cm²).

Advantages & Considerations

Advantages
  • Ready to use: no weighing, autoclaving or pouring
  • Target organisms identified by colony colour
  • Defined 25 cm² sampling area for comparable results
  • One format for surfaces, personnel and settle-plate air checks
Considerations
  • Some β-glucuronidase-negative E. coli strains, including most E. coli O157:H7, do not give the E. coli colour and are counted as coliforms.
  • A small number of non-coliform organisms may show β-galactosidase activity. Confirm atypical colonies with standard biochemical tests.
  • The medium does not contain disinfectant neutralisers (lecithin, polysorbate 80). Residual disinfectant on sampled surfaces may reduce recovery.
  • Contact plates suit flat, smooth surfaces only. Use swabs for irregular surfaces.

Related Standards (for user reference)

  • ISO 18593:2018 — Microbiology of the food chain — Horizontal methods for surface sampling
  • ISO 14698-1 — Cleanrooms and associated controlled environments — Biocontamination control
  • ISO 16649-2 — Enumeration of β-glucuronidase-positive Escherichia coli (TBX principle)

Frequently Asked Questions

What is the difference between a contact plate and a dip slide?
A contact plate is a 55 mm plate with a convex agar surface that you press onto a flat surface. It samples a defined 25 cm² area. Dip slides are dipped into liquids. AS-717 is supplied as a contact plate.

How do I sample a surface?
Remove the lid and press the agar surface firmly onto the surface for about 10 seconds with even pressure. Do not slide or rotate the plate. Replace the lid and incubate.

How do I read the result?
Blue-green colonies are E. coli. Magenta to purple colonies are other coliforms. Total coliforms are both colours added together. Report results as CFU per plate (CFU/25 cm²).

Do the plates contain a disinfectant neutraliser?
No. If you sample surfaces soon after disinfection, residual disinfectant may reduce recovery. Sample before cleaning or after the disinfectant has fully dried.

Technical Data Sheet — Coliform / E. coli Chromogenic Contact Plates

TDS-AS-717 · Rev 1.0 · November 2026
1. Intended Use

Ready-to-use 55 mm chromogenic contact plates for detecting and differentiating coliforms and E. coli on surfaces, on personnel and in air. For laboratory microbiological testing, quality control and research use only. Not intended for direct use in food, human or veterinary therapeutic applications, diagnosis, or consumption.

2. Mode of Action

SDS inhibits most Gram-positive organisms. β-D-galactosidase-positive coliforms cleave a chromogenic substrate (magenta/purple colonies). E. coli is detected by a second, E. coli-specific chromogenic substrate (blue-green colonies).

3. Formulation (g/L)
Component g/L
Meat peptone 10.00
Yeast extract 7.00
Sodium chloride 5.00
Sodium dodecyl sulfate (SDS) 0.20
Chromogenic substrates (proprietary blend) 0.12
Agar 14.50
Total 36.82
4. Product Specifications
Format Ready-to-use contact (RODAC-type) plates, 55 mm, convex agar surface
Fill volume 15.5 ± 1.0 mL per plate (convex meniscus above the rim)
Contact area Approximately 25 cm²
Pack size 10 plates per pack (as stated on label)
Storage 2–8 °C, in the dark, in the original sleeve, agar side up. Do not freeze.
Shelf life Expiry stated on label; batch-specific
pH (25 °C) 7.2 ± 0.2
5. Directions for Use
  1. Bring plates to room temperature. Discard any plate with condensation on the agar, contamination or damage.
  2. Remove the lid. Press the convex agar surface firmly onto the test surface for about 10 seconds with even pressure (about 500 g). Do not slide or rotate the plate.
  3. Personnel: press gloved fingertips gently onto the agar. Air: leave the open plate exposed, for example for 15 minutes.
  4. Replace the lid, label the plate, and clean the sampled area to remove residual medium.
  5. Incubate inverted as directed, then count colonies by colour.

Incubation: 35–37 °C, 18–24 hours (up to 48 hours if your procedure requires), aerobic.

6. Interpretation
Organism group Colony colour
Escherichia coli Blue-green
Other coliforms (Citrobacter, Enterobacter, Klebsiella, Serratia) Magenta / purple
Other Gram-negative organisms Colourless

Total coliforms = blue-green + magenta/purple colonies. Report as CFU per plate (CFU/25 cm²).

7. Quality Control
Organism Strain Inoculum (CFU) Growth Colony colour
Escherichia coli ATCC 25922 ≤100 Good growth Blue-green
Citrobacter freundii ATCC 8090 ≤100 Good growth Magenta / purple
Klebsiella aerogenes ATCC 13048 ≤100 Good growth Magenta / purple
Pseudomonas aeruginosa ATCC 27853 ≤100 Growth Colourless
Staphylococcus aureus ATCC 25923 10⁴–10⁶ Partial to complete inhibition Colourless, if any
8. Limitations
  • Some β-glucuronidase-negative E. coli strains, including most E. coli O157:H7, do not give the E. coli colour and are counted as coliforms.
  • A small number of non-coliform organisms may show β-galactosidase activity. Confirm atypical colonies with standard biochemical tests.
  • The medium does not contain disinfectant neutralisers (lecithin, polysorbate 80). Residual disinfectant on sampled surfaces may reduce recovery.
  • Contact plates suit flat, smooth surfaces only. Use swabs for irregular surfaces.
9. Disposal

Used plates contain viable microorganisms. Autoclave (121 °C, at least 15 minutes) or incinerate. If neither is possible, flood plates with 5–10 % bleach, leave overnight, then discard according to local regulations.

10. References
  1. Manafi M., Kneifel W., Bascomb S. (1991). Fluorogenic and chromogenic substrates used in bacterial diagnostics. Microbiol. Rev. 55:335–348.
  2. Manafi M. (2000). New developments in chromogenic and fluorogenic culture media. Int. J. Food Microbiol. 60:205–218.
  3. Frampton E.W., Restaino L., Blaszko N. (1988). Evaluation of the β-glucuronidase substrate X-GLUC in a 24-hour direct plating method for Escherichia coli. J. Food Prot. 51:402–404.
  4. Perry J.D. (2017). A decade of development of chromogenic culture media for clinical microbiology. Clin. Microbiol. Rev. 30:449–479.

Safety Data Summary — Coliform / E. coli Chromogenic Contact Plates

SDS-AS-717 · Rev 1.0 · November 2026 · Full 16-section SDS available on request
Hazard Classification

Not classified as hazardous according to the criteria of the GHS (Rev. 7) as adopted under the Australian WHS Regulations. No hazard pictogram or signal word is required for unused plates.

Biosafety

Used plates may contain viable coliforms and E. coli. Handle and dispose of them as biological waste.

Handling & Protection
  • Use good laboratory practice. Open plates only at the time of sampling.
  • Wear laboratory coat and gloves; safety glasses per laboratory policy.
First Aid
  • Skin: Wash with soap and water.
  • Eyes: Rinse with water for several minutes.
  • Ingestion: Rinse mouth. Seek medical advice if you feel unwell.
  • Used plates: After contact, wash thoroughly and follow your laboratory biosafety procedure.

Poisons Information Centre (Australia): 13 11 26 · Emergency: 000

Storage & Disposal

Store at 2–8 °C in the dark. Used plates contain viable microorganisms. Autoclave (121 °C, at least 15 minutes) or incinerate. If neither is possible, flood plates with 5–10 % bleach, leave overnight, then discard according to local regulations.

Transport

Unused plates are not dangerous goods. Transport at 2–8 °C where possible.

Quality & Certificate of Analysis

AS-717 · Batch-specific COA available on request

Each batch is manufactured in Melbourne and tested before release. A batch-specific Certificate of Analysis is available on request. Please quote the batch number printed on the sleeve.

Release Specifications
Test Specification
Appearance Clear to slightly opalescent gel, free of contamination and damage
Fill / meniscus Convex surface above the rim; 15.5 ± 1.0 mL
pH (25 °C) 7.2 ± 0.2
Sterility No growth after 48 h at 35–37 °C and 72 h at 20–25 °C (representative plates)
Growth performance Escherichia coli ATCC 25922: blue-green; Citrobacter freundii ATCC 8090: magenta / purple; Klebsiella aerogenes ATCC 13048: magenta / purple; Pseudomonas aeruginosa ATCC 27853: colourless
Selectivity Staphylococcus aureus ATCC 25923: partial to complete inhibition