AS-715

Salmonella Chromogenic Contact Plates 55 mm | AS-715

$79.00 AUD ex GST
Number
Request a Quote
Ready-to-use Contact Plates · Chromogenic

Salmonella Chromogenic Contact Plates

Ready-to-use 55 mm chromogenic contact plates for the presumptive detection of Salmonella on surfaces, with E. coli differentiation.

Cat. No. AS-71555 mm · 25 cm²Magenta SalmonellaReady to useRead at 24 h and 48 h
No preparation needed: plates are poured, sealed and ready to use. Press, incubate and read colonies by colour.

Colony Colours

SalmonellaMagenta / purple
E. coliBlue
Other organismsColourless / inhibited
Incubation35–37 °C

Intended Use

Ready-to-use 55 mm chromogenic contact plates for the presumptive detection of Salmonella on surfaces, with E. coli differentiation. For hygiene and environmental monitoring programs in food, pharmaceutical and industrial facilities.

For laboratory microbiological testing, quality control and research use only. Not intended for direct use in food, human or veterinary therapeutic applications, diagnosis, or consumption.

Principle

Peptones and yeast extract support growth, and selective agents suppress Gram-positive organisms and much of the accompanying Gram-negative flora.

Salmonella expresses C8-esterase, which cleaves a chromogenic substrate in the medium and gives magenta to purple colonies. E. coli cleaves a β-D-galactosidase substrate and forms blue colonies, so the two are easy to tell apart.

Composition

Component Function
Peptones and yeast extract Nitrogen, carbon, vitamins and growth factors
Inorganic salts Osmotic balance and buffering
Selective agents Inhibition of Gram-positive and part of the accompanying Gram-negative flora
C8-esterase chromogenic substrate Magenta / purple Salmonella colonies
β-D-galactosidase chromogenic substrate Blue E. coli colonies
Agar Gelling agent

Quantitative formulation is proprietary. Selective agents shown in amber.

Key Applications

Food ProcessingEnvironmental monitoring of food-contact and non-contact surfaces for Salmonella.
Poultry & EggHygiene checks in processing, hatchery and egg-handling areas.
Feed & IngredientsSurface monitoring in feed mills and dry-ingredient handling areas.
PharmaceuticalSurface monitoring in production and support areas.
Personnel & GlovesGlove-print and garment sampling for hygiene programs.
Air (Settle Plate)Open-plate exposure, for example 15 minutes, in production rooms.

Reading the Results

Organism group Colony colour
Salmonella spp. (presumptive) Magenta / purple
Escherichia coli Blue
Other organisms Colourless, or inhibited

Magenta colonies are presumptive Salmonella and must be confirmed by biochemical and serological tests before any decision is taken.

Advantages & Considerations

Advantages
  • Ready to use: no weighing, autoclaving or pouring
  • Target organisms identified by colony colour
  • Defined 25 cm² sampling area for comparable results
  • One format for surfaces, personnel and settle-plate air checks
Considerations
  • Results are presumptive. Some Proteus, Pseudomonas and other strains are C8-esterase positive and can form purple colonies. Confirm all magenta colonies.
  • Some Salmonella serovars may grow weakly or show weaker colour; re-examine negative plates at 48 hours.
  • The medium does not contain disinfectant neutralisers (lecithin, polysorbate 80). Residual disinfectant on sampled surfaces may reduce recovery.
  • Contact plates suit flat, smooth surfaces only. Use swabs and enrichment methods for irregular surfaces or low-level contamination.

Related Standards (for user reference)

  • ISO 18593:2018 — Microbiology of the food chain — Horizontal methods for surface sampling
  • ISO 6579-1 — Microbiology of the food chain — Detection of Salmonella spp. (reference method for confirmation)
  • ISO 14698-1 — Cleanrooms and associated controlled environments — Biocontamination control

Frequently Asked Questions

Is a magenta colony a confirmed Salmonella?
No. Magenta colonies are presumptive Salmonella. Some Proteus and Pseudomonas strains also give purple colonies. Confirm every magenta colony with biochemical and serological tests, for example by the ISO 6579-1 reference method.

How long should I incubate?
Incubate at 35–37 °C. Read at 24 hours and re-examine negative plates at 48 hours, because some strains develop colour slowly.

What colour is E. coli?
E. coli forms blue colonies, so it is easy to tell apart from magenta Salmonella.

Do the plates contain a disinfectant neutraliser?
No. If you sample surfaces soon after disinfection, residual disinfectant may reduce recovery. Sample before cleaning or after the disinfectant has fully dried.

Technical Data Sheet — Salmonella Chromogenic Contact Plates

TDS-AS-715 · Rev 1.0 · November 2026
1. Intended Use

Ready-to-use 55 mm chromogenic contact plates for the presumptive detection of Salmonella on surfaces, with E. coli differentiation. For laboratory microbiological testing, quality control and research use only. Not intended for direct use in food, human or veterinary therapeutic applications, diagnosis, or consumption.

2. Mode of Action

Selective agents suppress Gram-positive and much accompanying Gram-negative flora. C8-esterase-positive Salmonella cleave a chromogenic substrate (magenta/purple colonies); β-galactosidase-positive E. coli form blue colonies.

3. Composition
Component Function
Peptones and yeast extract Nitrogen, carbon, vitamins and growth factors
Inorganic salts Osmotic balance and buffering
Selective agents Inhibition of Gram-positive and part of the accompanying Gram-negative flora
C8-esterase chromogenic substrate Magenta / purple Salmonella colonies
β-D-galactosidase chromogenic substrate Blue E. coli colonies
Agar Gelling agent
4. Product Specifications
Format Ready-to-use contact (RODAC-type) plates, 55 mm, convex agar surface
Fill volume 15.5 ± 1.0 mL per plate (convex meniscus above the rim)
Contact area Approximately 25 cm²
Pack size 10 plates per pack (as stated on label)
Storage 2–8 °C, in the dark, in the original sleeve, agar side up. Do not freeze.
Shelf life Expiry stated on label; batch-specific
5. Directions for Use
  1. Bring plates to room temperature. Discard any plate with condensation on the agar, contamination or damage.
  2. Remove the lid. Press the convex agar surface firmly onto the test surface for about 10 seconds with even pressure (about 500 g). Do not slide or rotate the plate.
  3. Personnel: press gloved fingertips gently onto the agar. Air: leave the open plate exposed, for example for 15 minutes.
  4. Replace the lid, label the plate, and clean the sampled area to remove residual medium.
  5. Incubate inverted as directed, then count colonies by colour.

Incubation: 35–37 °C, aerobic. Read at 24 hours; re-examine negative plates at 48 hours.

6. Interpretation
Organism group Colony colour
Salmonella spp. (presumptive) Magenta / purple
Escherichia coli Blue
Other organisms Colourless, or inhibited

Magenta colonies are presumptive Salmonella and must be confirmed by biochemical and serological tests before any decision is taken.

7. Quality Control
Organism Strain Inoculum (CFU) Growth Colony colour
Salmonella Typhimurium ATCC 14028 ≤100 Good growth Magenta / purple
Salmonella Enteritidis ATCC 13076 ≤100 Good growth Magenta / purple
Escherichia coli ATCC 25922 ≤100 Good growth Blue
Staphylococcus aureus ATCC 25923 10⁴–10⁶ Inhibited —
Enterococcus faecalis ATCC 29212 10⁴–10⁶ Inhibited —
8. Limitations
  • Results are presumptive. Some Proteus, Pseudomonas and other strains are C8-esterase positive and can form purple colonies. Confirm all magenta colonies.
  • Some Salmonella serovars may grow weakly or show weaker colour; re-examine negative plates at 48 hours.
  • The medium does not contain disinfectant neutralisers (lecithin, polysorbate 80). Residual disinfectant on sampled surfaces may reduce recovery.
  • Contact plates suit flat, smooth surfaces only. Use swabs and enrichment methods for irregular surfaces or low-level contamination.
9. Disposal

Used plates contain viable microorganisms. Autoclave (121 °C, at least 15 minutes) or incinerate. If neither is possible, flood plates with 5–10 % bleach, leave overnight, then discard according to local regulations.

10. References
  1. Perry J.D., Ford M., Taylor J., Jones A.L., Freeman R., Gould F.K. (1999). ABC medium, a new chromogenic agar for selective isolation of Salmonella spp. J. Clin. Microbiol. 37:766–768.
  2. Manafi M. (2000). New developments in chromogenic and fluorogenic culture media. Int. J. Food Microbiol. 60:205–218.
  3. Gaillot O., di Camillo P., Berche P., Courcol R., Savage C. (1999). Comparison of CHROMagar Salmonella medium and Hektoen enteric agar for isolation of salmonellae from stool samples. J. Clin. Microbiol. 37:762–765.
  4. Perry J.D. (2017). A decade of development of chromogenic culture media for clinical microbiology. Clin. Microbiol. Rev. 30:449–479.

Safety Data Summary — Salmonella Chromogenic Contact Plates

SDS-AS-715 · Rev 1.0 · November 2026 · Full 16-section SDS available on request
Hazard Classification

Not classified as hazardous according to the criteria of the GHS (Rev. 7) as adopted under the Australian WHS Regulations. No hazard pictogram or signal word is required for unused plates.

Biosafety

Used plates may contain Salmonella (Risk Group 2). Handle used plates under PC2 conditions and decontaminate before disposal.

Handling & Protection
  • Use good laboratory practice. Open plates only at the time of sampling.
  • Wear laboratory coat and gloves; safety glasses per laboratory policy.
First Aid
  • Skin: Wash with soap and water.
  • Eyes: Rinse with water for several minutes.
  • Ingestion: Rinse mouth. Seek medical advice if you feel unwell.
  • Used plates: After contact, wash thoroughly and follow your laboratory biosafety procedure.

Poisons Information Centre (Australia): 13 11 26 · Emergency: 000

Storage & Disposal

Store at 2–8 °C in the dark. Used plates contain viable microorganisms. Autoclave (121 °C, at least 15 minutes) or incinerate. If neither is possible, flood plates with 5–10 % bleach, leave overnight, then discard according to local regulations.

Transport

Unused plates are not dangerous goods. Used plates or cultures sent for confirmation must be packed and shipped under infectious-substance rules (UN 3373 Category B, or UN 2814 where a Category A organism is suspected).

Quality & Certificate of Analysis

AS-715 · Batch-specific COA available on request

Each batch is manufactured in Melbourne and tested before release. A batch-specific Certificate of Analysis is available on request. Please quote the batch number printed on the sleeve.

Release Specifications
Test Specification
Appearance Clear to slightly opalescent gel, free of contamination and damage
Fill / meniscus Convex surface above the rim; 15.5 ± 1.0 mL
Sterility No growth after 48 h at 35–37 °C and 72 h at 20–25 °C (representative plates)
Growth performance Salmonella Typhimurium ATCC 14028: magenta / purple; Salmonella Enteritidis ATCC 13076: magenta / purple; Escherichia coli ATCC 25922: blue
Selectivity Staphylococcus aureus ATCC 25923: inhibited; Enterococcus faecalis ATCC 29212: inhibited

CUSTOMERS ALSO BOUGHT